If using the optional Waste Outlet Cap (60-5394-512 for 83B<br />

closures, 60-5394-513 for 100 mm closures), proceed with the<br />

following steps:<br />

3. Place the metal plate inside the plastic cap (Figure 2-2).<br />

The plate should be oriented so that the small tube will<br />

extend into the carboy. The small tube should extend 5 ±0.5<br />

cm long from nut. If the tube is missing, insert it now and<br />

finger-tighten the nut.<br />

Nut<br />

Metal Plate<br />

Spring-type<br />

Retainer<br />

5 cm tube extending<br />

into carboy<br />

Figure 2-2 Optional Waste Cap (83B cap shown)<br />

Place a retainer above and<br />

below the metal plate.<br />

4. The tubing is secured using metal, spring-type retainers. A<br />

pair of conventional pliers will allow you to open the<br />

retainer enough to slide it onto the tubing. On the larger<br />

diameter waste tube, slide one retainer about 2 inches (5<br />

cm) onto the tubing end.<br />

5. Feed this tube into the top of the metal plate until the<br />

retainer is against the top of the plate.<br />

6. Slide a retainer onto the tubing end until it is seated<br />

against the bottom of the metal plate.<br />

7. Thread the cap onto the carboy.<br />

8. The smaller diameter tube is the level sense tube. Push<br />

this tube about 2 cm into the top fitting and finger-tighten<br />

the nut.<br />