Section 2 Preparation<br />
CAUTION<br />
Elevated flammable vapor levels are possible. Ensure that the<br />
waste container is adequately ventilated, preferably by placing it in<br />
a fume hood.<br />
Note<br />
The Torrent is shipped with protective caps on the waste fittings.<br />
Remove these caps as you connect the waste lines. The internal<br />
solvent lines contain isopropyl alcohol, some of which may drain<br />
when the caps are removed.<br />
Note<br />
Waste connections on the left side panel are labeled<br />
with a Waste icon.<br />
The waste tube assembly is a bundle of two tubes that routes<br />
post-run valve wash fluid to a waste container. Instructions for<br />
diverter valve waste tubing is covered in the INSTALL<br />
OPTIONAL EQUIPMENT section (2.10).<br />
To connect the waste tube assembly to the Torrent:<br />
1. Select the tube with the pre-swaged nut and ferrule and<br />
connect this end to the port labeled VALVE. The connection<br />
should be finger-tight. Then, use a wrench to tighten the<br />
fitting an additional 1 8 turn.<br />
2. Connect the other line to the LEVEL port:<br />
a. Locate a nut (60-0923-015) and ferrule (60-0083-163) in<br />
the accessory package (60-5249-002).<br />
b. Insert the ferrule, narrow end first, into the LEVEL<br />
port.<br />
c. Loosely thread the nut onto the LEVEL port.<br />
d. Insert the waste air line into the LEVEL port until it is<br />
fully seated.<br />
e. Finger-tighten the nut. If desired, tighten the nut an<br />
additional 1 2 turn using a wrench.<br />
The other ends of the Waste line should be placed in the waste<br />
container.<br />
2-9