2.6.3 Pressurized Solvent Lines in the Laboratory<br />
When receiving solvent from pressurized lines on-site, the<br />
Torrent’s solvent level sensing feature is unsupported (and<br />
unnecessary).<br />
Note<br />
The solvent pressure should be regulated to not exceed 5 psi.<br />
To complete the solvent connections:<br />
1. Ensure that the solvent supply lines have shut-off valves.<br />
These valves must be accessible to the user.<br />
Note<br />
The solvent supply should be shut-off when the system is not in<br />
use. Otherwise, pressurized solvent may be forced through the<br />
<strong>CombiFlash</strong> Torrent pump seals and leak from the waste port or<br />
other fluid path from the automated injection valve.<br />
2. Locate the 1 8-inch plug, part number 209-0165-21, in<br />
accessory kit 60-5249-002. Install this plug on the B Prime<br />
port.<br />
3. Connect user-supplied tubing with 3 8-inch fittings from the<br />
source to the A Inlet and B Inlet ports on the side panel.<br />
Note<br />
Solvent Level Sensing and B Solvent Prime options should be<br />
disabled when completing the system configuration settings in<br />
section 2.13.1.<br />
2.7 Connect Waste Lines<br />
WARNING<br />
Risk of fire or equipment damage. Failure to connect Waste<br />
Port tubing may allow organic solvents to pool in unsafe areas,<br />
possibly creating dangerous levels of flammable vapors.<br />
2-8