2.6.3 Pressurized Solvent Lines in the Laboratory<br />

When receiving solvent from pressurized lines on-site, the<br />

Torrent’s solvent level sensing feature is unsupported (and<br />

unnecessary).<br />

Note<br />

The solvent pressure should be regulated to not exceed 5 psi.<br />

To complete the solvent connections:<br />

1. Ensure that the solvent supply lines have shut-off valves.<br />

These valves must be accessible to the user.<br />

Note<br />

The solvent supply should be shut-off when the system is not in<br />

use. Otherwise, pressurized solvent may be forced through the<br />

<strong>CombiFlash</strong> Torrent pump seals and leak from the waste port or<br />

other fluid path from the automated injection valve.<br />

2. Locate the 1 8-inch plug, part number 209-0165-21, in<br />

accessory kit 60-5249-002. Install this plug on the B Prime<br />

port.<br />

3. Connect user-supplied tubing with 3 8-inch fittings from the<br />

source to the A Inlet and B Inlet ports on the side panel.<br />

Note<br />

Solvent Level Sensing and B Solvent Prime options should be<br />

disabled when completing the system configuration settings in<br />

section 2.13.1.<br />

2.7 Connect Waste Lines<br />

WARNING<br />

Risk of fire or equipment damage. Failure to connect Waste<br />

Port tubing may allow organic solvents to pool in unsafe areas,<br />

possibly creating dangerous levels of flammable vapors.<br />