Section 6 Maintenance<br />

6.6 Electrical Troubleshooting<br />

If your instrument stops working and the touch panel display<br />

is off, check the line cord connection.<br />

If the line cord is connected properly, check the fuses in the<br />

power input module located on the system’s rear panel. These<br />

are time delay fuses (T-type) and rated according to the mains<br />

input power voltage:<br />

• 100 to 129 VAC mains input: 3.0 A fuse.<br />

• 215 to 264 VAC mains input: 1.5 A fuse.<br />

CAUTION<br />

Always replace fuses with the same type and rating.<br />