Note<br />

Pumping solvent B at a moderate flow rate (150 to 200 mL/min)<br />

over time will usually solubilize obstructions. Generally, the<br />

recommended solvent is the highest polarity solvent you have<br />

recently used (solvent B).<br />

8. Monitor the Raw Lamp Energy gauge. As the system<br />

pumps solvent, the raw lamp energy should gradually<br />

improve. Pump solvent for two to five minutes or until the<br />

indicator reaches the far-right of the green range.<br />

When the indicator is in the yellow, the lamp energy and flow<br />

cell are ready for the next run. However, continuing the<br />

cleaning for greater raw lamp energy should reduce the<br />

frequency of flow cell cleaning messages (Figure 6-2).<br />

If after five minutes the indicator is still in the red range,<br />

repeat the cleaning steps using an alternative solvent<br />

(Table 6-1). Or, complete the steps in the Monthly Flow Cell<br />

Cleaning procedure (section 6.5.3).<br />

Note<br />

If the numerical values of the Raw Lamp Energy (Figure 6-3) do<br />

not change, or if the first number remains at zero, contact<br />

Teledyne <strong>Isco</strong>’s Technical Service department.<br />