Pumping solvent B at a moderate flow rate (150 to 200 mL/min)<br />
over time will usually solubilize obstructions. Generally, the<br />
recommended solvent is the highest polarity solvent you have<br />
recently used (solvent B).<br />
8. Monitor the Raw Lamp Energy gauge. As the system<br />
pumps solvent, the raw lamp energy should gradually<br />
improve. Pump solvent for two to five minutes or until the<br />
indicator reaches the far-right of the green range.<br />
When the indicator is in the yellow, the lamp energy and flow<br />
cell are ready for the next run. However, continuing the<br />
cleaning for greater raw lamp energy should reduce the<br />
frequency of flow cell cleaning messages (Figure 6-2).<br />
If after five minutes the indicator is still in the red range,<br />
repeat the cleaning steps using an alternative solvent<br />
(Table 6-1). Or, complete the steps in the Monthly Flow Cell<br />
Cleaning procedure (section 6.5.3).<br />
Note<br />
If the numerical values of the Raw Lamp Energy (Figure 6-3) do<br />
not change, or if the first number remains at zero, contact<br />
Teledyne <strong>Isco</strong>’s Technical Service department.<br />
6-8