07.03.2014 Views

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 6 Maintenance<br />

• Red – lamp energy is obstructed to a degree that the<br />

system might not reliably detect peaks. If you<br />

attempt to operate the system, peak collection will be<br />

forced to collect all. This prevents diverting desired<br />

compounds to waste.<br />

• Yellow – lamp energy is partially obstructed and the<br />

peak detector performance might be impaired. The<br />

system can be operated although the absorbance<br />

trace might appear noisy. Yellow might be a normal<br />

indication for systems with flow cell windows stained<br />

by compounds.<br />

• Green – lamp energy is sufficient to detect peaks up<br />

to 2.4 Absorbance Units.<br />

3. Remove the column and insert a bypass tube (P/N<br />

209-0165-76) between the upper and lower column mounts.<br />

4. Set the Flow Rate to 200 mL/min (Figure 6-4).<br />

5. Select Through column for the Valve Position option.<br />

6. Select Next for the Pump into Tube # option. By pumping<br />

into a collection tube, the diverter valve is also cleaned of<br />

any residue during this operation.<br />

7. Click the Prime B button to pump 100% Solvent B through<br />

the bypass tube and into the collection tubes.<br />

Figure 6-4 <strong>Manual</strong> Control Settings - Quick Cleaning<br />

6-7

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!