<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

CAUTION<br />

When using methanol as solvent B with silica column media, do<br />

not pump more than 50% solvent B. Higher percentages of<br />

methanol might break down the silica structure, possibly causing<br />

obstructions in the flow path.<br />

6.5.2 Quick Cleaning when Recommended<br />

When the lamp energy is lower than normal, the system will<br />

recommend flow cell cleaning (Figure 6-2) before starting a<br />

separation run.<br />

When the system displays this message you can:<br />

• Cancel Run (recommended) – Click the Cancel Run<br />

button so you can perform a quick cleaning described in<br />

the following steps.<br />

• Continue Collect All – Click this button to ignore the<br />

message. Because the peak detection operation might be<br />

impaired, the system automatically collects all fluids to<br />

avoid diverting compounds of interest to waste.<br />

• Help – Click this button to display the flow cell cleaning<br />

on-line help topic.<br />

To perform a quick cleaning:<br />

1. After clicking the Cancel Run button, select the<br />

Tools><strong>Manual</strong> Control menu option.<br />

2. From the <strong>Manual</strong> Control window (Figure 6-3), note the<br />

Raw Lamp Energy level at 254 nm. The Raw Lamp Energy<br />

gauge has three ranges: red, yellow, and green.<br />

Figure 6-3 Raw Lamp Energy Gauge<br />

6-6