Section 6 Maintenance<br />

End of<br />

Separation<br />

Flush<br />

100<br />

Separation<br />

Gradient<br />

Initial %B Next Run Initial %B<br />

End of<br />

Run<br />

75<br />

50<br />

25<br />

0<br />

Gradient % Solvent B<br />

Time or Column Volumes<br />

Figure 6-1 Default post-separation column and flow cell flush<br />

Figure 6-2 Flow cell cleaning message<br />

Typically, chemists Stop and then Fast Forward the run after<br />

the last compound elutes. This action skips the<br />

post-separation flush. If any of the above conditions appear,<br />

consider allowing some of the runs to continue through the<br />

flush, or run a high percentage of %B solvent through the<br />

system for a few minutes at the end of each day.<br />

If the separation runs always continue through the flush and<br />

the conditions still occur frequently, edit the default column<br />

methods to extend the flush duration.<br />