6.2 Preventive Maintenance<br />
The system requires preventive maintenance by the user for<br />
safe and reliable operation. Refer to the schedule below for the<br />
minimum periodic maintenance requirements.<br />
As Needed – Perform these tasks as conditions require:<br />
• Cleaning (section 6.3).<br />
• Quick flow cell cleaning when recommended by a system<br />
alert message (section 6.5.2).<br />
Every Run – Allow the separation run to finish with a high<br />
percentage of solvent B to flush residual compounds from the<br />
column, internal tubing, and flow cell. Refer to section 6.5.1.<br />
Monthly – Perform these tasks at least monthly, more<br />
frequently if conditions warrant:<br />
• Tubing Inspection (section 6.4).<br />
• Collection rack and tray cleaning (section 6.3.1).<br />
• Monthly flow cell cleaning (section 6.5.3).<br />
Additional periodic maintenance tasks should be performed at<br />
one and five year intervals by a qualified technician. These<br />
tasks include:<br />
Yearly – Perform the tasks identified in Teledyne <strong>Isco</strong><br />
instruction sheet 60-5243-184.<br />
Every Five Years – Replace internal battery.<br />
Contact Teledyne <strong>Isco</strong>’s Technical Service Department for<br />
more information regarding periodic maintenance.<br />
6-2