CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />
16.Gas Inlet – The Torrent has an internal air pump that<br />
delivers air through the column for a post-run purge. The<br />
air is also used to detect solvent and waste levels. If you<br />
desire to use a specific gas such as nitrogen, connect the<br />
gas source to this port.<br />
17.Pressure Reference – When detecting solvent or waste<br />
levels in containers or conduits not exposed to atmospheric<br />
pressure, a headspace reference pressure is required so<br />
that the Torrent can read the differential pressure to<br />
determine levels.<br />
18.A Inlet – This port receives the A solvent.<br />
19.A Level – This port measures the solvent A level by means<br />
of pressurized air or gas source.<br />
20.B Prime – This port primes the B solvent pump.<br />
21.B Inlet – This port receives the B solvent.<br />
22.B Level – This port measures the solvent B level by means<br />
of pressurized air or gas source.<br />
23.Waste Valve – This port delivers waste fluids to a<br />
user-supplied waste container.<br />
24.Waste Level – The Torrent uses air or user-supplied gas to<br />
measure the hydrostatic pressure of the waste above the<br />
level sense line end. The system will suspend operation<br />
when this user-set level is too high.<br />
25.Fraction Valve – The Torrent controls the optional<br />
Fractionation Valve through this port.<br />
26.Sample Load Pump – The Torrent controls the optional<br />
Sample Load Pump through this port.<br />
27.Mains Power – Connects the Torrent to AC line voltage.<br />
28.Power Outlet – Provides AC line voltage to an external<br />
device that meets the 115 VAC and 2.0 ampere or 230 VAC<br />
and 1.0 ampere rating.<br />
CAUTION<br />
The back panel power connector is intended only for use with the<br />
Foxy R2 fraction collector. Do not connect other devices to this<br />
outlet.<br />
1-8