Section 4 <strong>User</strong> Help Reference<br />

TLC, wait at least the time established as T. After this time<br />

has elapsed it is likely that the desired compound has come<br />

off the cartridge and you can safely click the Cartridge<br />

Bypass button.<br />

For example, a TLC plate run at 30%B revealed that the last<br />

desired compound had an Rf factor of 0.35. An empty 25-gram<br />

size solid load cartridge was filled with sample adsorbed onto<br />

10 grams of silica, and then placed on the system. The<br />

programmed flow rate was 60 mL/min. During the run, the<br />

flow rate dropped to 30 mL/min. The equation is then:<br />

T = 45 mL / (0.35 * 30 mL/min) = 4.3 minutes<br />

In this example, the Cartridge Bypass feature was used 4.3<br />

minutes after the solvent gradient reached 30% B.<br />

The calculation is conservative, but allows you to minimize the<br />

time needed to keep the solid load cartridge in the fluid path.<br />

4.4.9 Viewing runs<br />

After completing a run, the PeakTrak Main window is used to<br />

display all collected run data. You can also open previous runs<br />

to view the chromatogram and the peak/tube locations. To<br />

open a previous run:<br />

1. From the main window, select File>Open, or click the Open<br />

button. The FILES [4.3.5] window is displayed.<br />

2.Choose a Run file and click the Open button.<br />

3.PeakTrak displays the run in the MAIN WINDOW [4.3.1].<br />

The PeakTrak Main window will include:<br />

• Rack and tube information — The left pane of the<br />

window lists the current rack, a map for that rack, and a<br />

table that lists the peaks and their corresponding tube<br />

numbers. If the window is currently displaying the<br />

collection parameters, click the Rack View button to<br />

view this information.<br />

• Method parameters — Click the Display Method<br />

button to view a summary of the peak detection and<br />

collection settings for the run.<br />