07.03.2014 Views

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

CombiFlash Torrent® User Manual - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Section 4 <strong>User</strong> Help Reference<br />

4.4.7 Monitoring the Wavelength Ratio<br />

When using two peak detection options, such as 1 with 2 or<br />

All Wavelength Detection with External, a ratio of the<br />

detection sources can also be displayed, which at times can<br />

provide the best indication of compound purity. Refer to the<br />

following discussion.<br />

If a pure compound is eluting, the absorbance is linearly<br />

related to the concentration of the compound in the solvent. If<br />

the compound absorbs differently at different wavelengths,<br />

the absorbance at each wavelength may be different, but still<br />

linearly related. For example, assume a compound eluting<br />

from the module has an absorbance equal to 2 times the<br />

concentration at 254 nm. This same compound at 220 nm has<br />

an absorbance of 1.5 times the concentration. The ratio of<br />

these signals will be 1.33. Since the relationship of absorbance<br />

to concentration is not variable, the ratio remains steady while<br />

the concentration changes from the beginning to the end of the<br />

peak. During the duration of the peak, the ratio will be 1.33<br />

and this constant value is displayed as a horizontal line.<br />

Now assume a case where there is a second compound eluting,<br />

only slightly shifted in time from the original compound. It is<br />

possible that the detection absorbance trace alone would<br />

indicate a single, valid chromatographic peak. In reality, it is a<br />

combination of two peaks. By monitoring a second wavelength,<br />

it may be possible to reveal the second compound. Because of<br />

the slight shift in time and the different absorbance properties<br />

of the two compounds, the changing ratio during the detected<br />

peak would reveal the impurity. Therefore, one can therefore<br />

assume that if the ratio is not constant for the entire duration<br />

of the peak, the compound eluting may not be pure.<br />

To display the ratio, open the METHOD EDITOR [4.3.2] window<br />

and select the Show Ratio option.<br />

4.4.8 Bypassing the Solid Load Cartridge<br />

Occasionally, impurities precipitate in the solid load cartridge<br />

as compounds are purified. This may cause high back<br />

pressures resulting in reduced flow rates and long run times.<br />

Clicking this button changes the injection valve position,<br />

which removes the solid load cartridge from the solvent path<br />

and relieves the back pressure.<br />

4-45

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!