Section 4 <strong>User</strong> Help Reference<br />
save the method using the FILE>Save or Save As menu<br />
commands.<br />
4.4.3.3 Scaling Method Files to Different Column Sizes<br />
If you want to scale a method file to a different column size,<br />
use the FILE>Save As menu command. From the FILES [4.3.5]<br />
window, select the Scale option, then choose the new column<br />
size and click the Save button. The module will create a new<br />
method file with the gradient and run parameters scaled to<br />
the new column size.<br />
This feature is useful when developing methods using small<br />
sample volumes. After you optimize the method to successfully<br />
extract the compounds of interest, you can then scale up the<br />
method to load and extract larger volumes using larger<br />
RediSep column sizes.<br />
4.4.4 Editing a Default Method<br />
Default Methods allow you to quickly recall standard method<br />
settings. There are standard methods for each size and type of<br />
RediSep column. When shipped from the factory, the Default<br />
Methods were set to the settings recommended by Teledyne<br />
<strong>Isco</strong>.<br />
The Default Method settings can be tailored to your<br />
application. To modify the Default Methods:<br />
1. Select the TOOLS>EDIT DEFAULTS [4.2.3.5] menu command.<br />
2.If prompted for a password, enter the module password.<br />
(The SET PASSWORD [4.3.11] window controls the password<br />
protection feature.)<br />
3.Select the column size and media from the RediSep Column<br />
list.<br />
4.Change the settings to meet your specific application<br />
requirements. Refer to the EDITING A METHOD [4.4.2]<br />
instructions.<br />
5.Click the Save button (or File>Save) to replace the default<br />
method settings.<br />
You can reset the method to Teledyne <strong>Isco</strong>-recommended<br />
defaults. While viewing the default method file for a selected<br />
column size and type, click the “Reset file to <strong>Isco</strong> defaults”<br />
4-43