Section 4 <strong>User</strong> Help Reference<br />

• RediSep Rf 3 kg = 4.8 L<br />

Other media types will have similar column volumes. Refer to<br />

the REDISEP RF COLUMN SELECTION GUIDE [ON-LINE] for actual<br />

volumes.<br />

As an example, consider a RediSep Rf 120 g column and a 3<br />

column volumes setting. Before instructing you to inject the<br />

sample, the module will pump 576 mL of solvent A through<br />

the column and out the waste port.<br />

Initial Waste & Time Windows The initial waste diverts a<br />

user-determined line volume to the waste port until the eluant<br />

is about to be collected. At the start of a run, the internal<br />

solvent lines hold some solvent from the previous run. This<br />

volume can be diverted to waste, along with the anticipated<br />

volume of fluid that will pass through the column before a<br />

compound of interest will elute.<br />

Time Windows can limit the fraction collection to specific time<br />

durations of the run. To use time windows enter a start time<br />

and an end time. All times are relative to start of the run, just<br />

following the sample injection. Then choose a peak collection<br />

mode. If you select Peaks the module will collect fluid only<br />

when the time window is active and a peak is present. If you<br />

select All the module will collect all fluid during the time<br />

window regardless of the peak state. Define up to three time<br />

windows.<br />

Air Purge You can program the module to purge the column<br />

with air after a run. This feature allows you to remove<br />

solvents from the column before it is discarded. Enter the<br />

desired time to purge the column, in minutes. To disable the<br />

post-run air purge, enter zero minutes.<br />

Peak Collection There are three peak collection options: All,<br />

Peaks, or None. The All option will collect peak and non-peak<br />

fluid in tubes. The Peaks option will collect only peak fluid in<br />

tubes and divert all other fluids to the waste port. The None<br />

option will divert all fluids, peak and non-peak, to the waste<br />

port. The None option is useful for developing custom methods<br />

that perform a column wash, system cleaning, or similar<br />

function where solvents and elute do not need to be collected.<br />

4-39