CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
CombiFlash Torrent® User Manual - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Section 4 <strong>User</strong> Help Reference<br />
If you will transfer the sample mixture directly from a<br />
container, attach the tubing to the pump inlet. Next,<br />
place the weighted end into the mixture container.<br />
c. Place the On/Off (I/O) switch in the On position.<br />
Adjust the pump speed using the control knob.<br />
d. When the desired sample volume has been loaded,<br />
place the On/Off (I/O) switch in the Off position.<br />
e. Some of the sample mixture will remain in the tubing<br />
before the automated injection valve. In most<br />
applications, pump a chase solution to force the<br />
sample mixture through the injection valve. To do so,<br />
repeat steps b through d using a suitable chase<br />
solvent.<br />
You may allow the module to proceed with the run while<br />
monitoring the progress on the PeakTrak Main window. If<br />
desired, gradient parameters may be modified during the run.<br />
(See REAL-TIME GRADIENT EDITING [4.4.6].)<br />
As the run progresses, the absorbance trace is drawn on the<br />
Gradient Plot area. The purification continues until the end of<br />
the run defined by the run length setting.<br />
Note: Solvent may be under pressure. Do not remove a<br />
column or solid load cartridge until the pressure gauge on the<br />
touch screen to return to 0 psi.<br />
Note: After use, pump clean solvent to remove sample<br />
residue.<br />
See Also:<br />
REDISEP RF COLUMN SELECTION GUIDE [ON-LINE]<br />
SOLVENT SELECTION TABLE [ON-LINE]<br />
EDITING A METHOD [4.4.2]<br />
EDITING A DEFAULT METHOD [4.4.4]<br />
BYPASSING THE SOLID LOAD CARTRIDGE [4.4.8]<br />
4.4.2 Editing a method<br />
Editing a method allows you to tailor the operation of the<br />
<strong>CombiFlash</strong> Torrent module to best separate or purify the<br />
compounds of interest.<br />
PeakTrak has two types of method settings — basic and<br />
advanced.<br />
4-37