<strong>CombiFlash</strong> Torrent® <strong>User</strong> <strong>Manual</strong><br />

If you selected Solid (Pause), the module equilibrates<br />

the column with the starting %B and then waits while<br />

you prepare the solid sample load cartridge. Place the<br />

cartridge on the module and click OK to continue.<br />

If you selected Solid, the module equilibrates the<br />

column and immediately starts the separation.<br />

If you selected None (on column), the module skips the<br />

equilibration and immediately starts the separation.<br />

If you selected Liquid, the module equilibrates the<br />

column and then waits. When instructed, perform either<br />

a Syringe Injection or use the Sample Load Pump:<br />

Syringe Injection:<br />

a. Connect the Liquid Injection Accessory from the<br />

accessory kit to the Sample Injection Port.<br />

b. Place the syringe on top of the Liquid Injection<br />

Accessory valve.<br />

c. Open the valve by turning the handle so it is in line<br />

with the tubing.<br />

d. Inject the liquid sample mixture.<br />

e. Close the valve by turning the handle so that it is<br />

perpendicular to the tubing. Remove the syringe.<br />

f. Inject a chase solvent by connecting a syringe loaded<br />

with a chase solvent to the Liquid Injection<br />

accessory’s luer fitting.<br />

g. Open the valve by turning the handle so it is in line<br />

with the tubing.<br />

h. Inject at least 10 mL of the chase solvent.<br />

i. Close the valve by turning the handle so that it is<br />

perpendicular to the tubing. Remove the syringe.<br />

j. Click the OK button to continue the run.<br />

Sample Load Pump:<br />

a. Connect the outlet tubing from the Sample Load<br />

pump to the Torrent module’s Sample Injection Port.<br />

b. Prepare the Sample Load pump to receive the sample:<br />

If the sample mixture will be transferred to the pump,<br />

insert the empty cartridge body onto the pump inlet.<br />

Then, fill the cartridge with a measured mixture<br />

volume.<br />

4-36