Section 4 <strong>User</strong> Help Reference<br />

by selecting the TOOLS>MANUAL CONTROL [4.2.3.3] menu<br />

command.<br />

The configuration window displays four tabs: INSTRUMENT<br />

CONFIGURATION [4.3.10.1], NETWORK CONFIGURATION [4.3.10.2],<br />

and USER PREFERENCES [4.3.10.3], FRACTION COLLECTOR<br />

[4.3.10.4].<br />

4.3.10.1 Instrument Configuration<br />

Solvent Controls<br />

• Add Solvent — Click this button to enter a solvent<br />

name. PeakTrak can store up to 16 solvent names. For<br />

assistance in selecting appropriate solvents, see the<br />

SOLVENT SELECTION TABLE [ON-LINE].<br />

• Remove Solvent — Click this button to open a list of<br />

all entered solvent names. Select the names you want to<br />

remove and then click the OK button.<br />

• Enable Solvent Level Sensing — If this option is<br />

selected, the module will check the solvent levels before<br />

and during a run. PeakTrak will alert you if a solvent<br />

level becomes too low. Note that the solvent level<br />

monitoring option does not control the operation of<br />

Waste Solvent Level monitoring. Waste level sensing is<br />

always enabled at the beginning of a run.<br />

• Minimum Solvent Level — This value is<br />

density-dependent, thus a range is given. To understand<br />

the range, consider that dichloromethane is more dense<br />

than hexane. If the selected range is 1–4, an alert would<br />

be triggered when the dichloromethane falls to about 1<br />

cm above the bottom of the weighted filter. At the same<br />

selected range, an alert would be triggered when the<br />

hexane solvent level falls to about 4 cm above the<br />

weighted filter.<br />

• Waste Level Sensing — The module can detect full<br />

container conditions if the air line is placed in the waste<br />

collection container. If waste level detection is desired<br />

enable the waste level sensing feature (recommended).<br />

Otherwise, uncheck this option to disable this feature.<br />