Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />
Section 5 Maintenance<br />
Figure 5-4 Internal Desiccant<br />
9. Replace the front panel using care to ensure that the wiring<br />
will be free of the gear train.<br />
10. Replace the bezel and screws. Tighten the 10 screws in an<br />
even, cross-torquing pattern.<br />
The internal humidity indicator should return to its normal blue<br />
color in a few hours.<br />
5.7.1 Renewing the<br />
Desiccant<br />
CAUTION<br />
Desiccant may produce irritating fumes when heated. To<br />
reduce the hazard of the fumes:<br />
• Use a vented oven in a well-ventilated room<br />
• Do not remain in the room while recharging is taking place<br />
• Use the recommended temperature.<br />
There have been reports of irritating fumes coming from the<br />
desiccant during reactivation. While our attempts to duplicate<br />
the problem have been unsuccessful, we still urge you to use<br />
caution. Material Safety Data Sheets are provided in<br />
Appendix C.<br />
To renew the desiccant:<br />
1. Remove the bag from the <strong>Glacier</strong> controller.<br />
2. Place a sheet of brown paper on a flat metal sheet. You can<br />
use a brown grocery bag and a typical cookie sheet.<br />
3. Place the bag on the sheet. If you are recharging several<br />
bags, do not stack the bags on top of each other or allow<br />
them to touch.<br />
4. Place in a vented, circulating forced air, convection oven in<br />
a well-ventilated room. Allow two inches of air space<br />
between the top of the bag and the next rack. Keep the tray<br />
a minimum of 16 inches from the heating element.<br />
5-9