Section 2 Installation<br />
TAKING 200 ml<br />
CALIBRATE SAMPLE<br />
The <strong>Glacier</strong> displays the screen at left and goes through a complete<br />
sample collection cycle. The <strong>Glacier</strong> deposits the sample in<br />
the graduated cylinder.<br />
Note<br />
The volume delivered during calibration is the programmed<br />
Sample Volume.<br />
2.10.6 Calibration Step 6 The <strong>Glacier</strong> displays the amount of liquid it has deposited into<br />
the graduated cylinder.<br />
VOLUME DELIVERED<br />
200 ml<br />
Measure the volume and compare the numbers. If they match,<br />
press ENTER and the <strong>Glacier</strong> will return to the Standby state.<br />
Note<br />
Place the graduated cylinder on a level surface. Read the liquid<br />
volume from the bottom of the meniscus, not the top of the<br />
curved edges where the liquid wets the sides of the cylinder.<br />
If the amounts differ, enter the actual volume delivered. To enter<br />
the actual volume:<br />
1. Press the appropriate NUMBER buttons on the keypad.<br />
Note<br />
If you enter an incorrect value with the number buttons, press<br />
the STOP button. The <strong>Glacier</strong> restores the original value and<br />
waits for a new value.<br />
___ ml! ARE YOU<br />
SURE? YES NO<br />
2. Press the Enter button to accept the value. The <strong>Glacier</strong><br />
updates its internal pump tables. If there is a significant<br />
difference between the old and new values, the <strong>Glacier</strong> displays<br />
the screen at left.<br />
3. Compare this value with the measured volume.<br />
If they match, select YES using the ARROW button and<br />
then press ENTER.<br />
If they do not match, select NO using the ARROW button<br />
and then press ENTER. The <strong>Glacier</strong> returns to the Volume<br />
Delivered screen (step 1, above).<br />
When through with Calibration Step 6, the <strong>Glacier</strong> logs the calibration<br />
time in the event log and returns to the Standby state. Be<br />
sure to reconnect the pump tube end to the bulkhead fitting.<br />
2.11 Start the Program The <strong>Glacier</strong> has two programs—the current program and the<br />
stored program. To run the current program press the GO<br />
button.<br />
1 2 3<br />
You can also quickly recall the stored program and start it. To do<br />
so, press the PROGRAM button twice, followed by the GO button.<br />
This action will load the stored program settings as the current<br />
settings and run the program.<br />
2-18