To enter the Suction Head:<br />
1. Measure the Suction Head height (Figure 2-9).<br />
2. Enter the height. Press the appropriate NUMBER buttons<br />
on the keypad.<br />
3. Press the ENTER button to accept the value. The <strong>Glacier</strong><br />
loads the fixed Suction Head into the current program settings<br />
and advances to the next step.<br />
Figure 2-9 Measuring Suction Head Height<br />
2.10.4 Calibration Step 4 The <strong>Glacier</strong> advances to the Check Sample Volume display.<br />
CHECK SAMPLE VOLUME?<br />
YES NO<br />
Checking the sample volume is recommended if ultimate<br />
accuracy is required by your sampling protocol.<br />
The default selection is NO. To accept this, press the ENTER<br />
button. The <strong>Glacier</strong> will return to the Standby state.<br />
To check sample volumes, select YES by pressing the ARROW<br />
button, and then press the ENTER button. The <strong>Glacier</strong> will continue<br />
with the calibration steps.<br />
2.10.5 Calibration Step 5 If YES was selected in Calibration Step 4, the sampler waits<br />
while you prepare to collect a sample.<br />
CALIBRATE VOLUME<br />
PRESS WHEN READY!<br />
To prepare for a sample:<br />
1. Pull the lower pump tube end from the bulkhead fitting.<br />
2. Hold the pump tube outlet over a graduated cylinder, such<br />
as Teledyne <strong>Isco</strong>’s 1000-ml plastic graduated cylinder, P/N<br />
299-0020-00.<br />
3. Press the ENTER button and the <strong>Glacier</strong> will start to collect<br />
the sample.<br />
2-17