07.03.2014 Views

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />

Section 2 Installation<br />

1.<br />

2.<br />

Figure 2-5 Attaching the suction line to the pump tubing<br />

Once the coupler is attached to the pump tube, removal is difficult,<br />

and may require cutting the tube.<br />

2.8.3 Connecting PTFE<br />

Suction Line<br />

To connect the 3 /8-inch (9-mm) PTFE suction line to the pump<br />

tube:<br />

Items required:<br />

Suction line with strainer attached<br />

3 /4-inch (19-mm) diameter hose clamp (plastic or<br />

stainless steel recommended)<br />

1. Place a hose clamp on the upper pump tube.<br />

2. Insert about 1 inch (25 mm) of the PTFE suction line into<br />

the upper pump tube.<br />

3. Position the clamp over the joined area and tighten it.<br />

2.8.4 Strainers The 3 /8 inch ID vinyl suction lines are shipped from the factory<br />

with our standard weighted polypropylene strainer (Figure 2-6)<br />

installed on one end of the suction line and a tubing coupling on<br />

the other end.<br />

Additionally, Teledyne <strong>Isco</strong> offers two low flow stainless steel<br />

strainers (Figure 2-7) for 1 /4 inch ID and 3 /8 inch ID suction lines.<br />

To install the low flow strainer in PTFE suction line, first heat<br />

the end of the suction line to make it more pliable, then carefully<br />

screw the threaded end of the strainer into the suction line.<br />