07.03.2014 Views

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />

Section 2 Installation<br />

CAUTION<br />

Never use a DC extension cable without first consulting with a<br />

Teledyne <strong>Isco</strong> Service Technician. They will advise you of the<br />

proper wire gauge for the length you require.<br />

The refrigerator compressor is equipped with a battery controller<br />

that will cut out to protect the compressor and the battery when<br />

the voltage drops to 10.4–11.3 VDC. The controller will cut in at<br />

11.7–12.5 VDC.<br />

CAUTION<br />

Never charge the battery while connected to the <strong>Glacier</strong>.<br />

Over-voltages could damage the electronics.<br />

Note<br />

The refrigerator housing has a High/Low power switch<br />

mounted near the DC input power connector. This High/Low<br />

switch is unused. The switch must remain in the Low position<br />

for normal operation.<br />

2.7.2 Battery<br />

Recommendations<br />

2.8 Connect a Suction<br />

Line and Strainer<br />

Before each sampling program, the battery should be exchanged<br />

with a fully-charged battery.<br />

Power consumption is mostly determined by the ambient temperature.<br />

Higher ambient temperatures will in turn create a higher<br />

demand for power. A battery selection guide is provided in<br />

Appendix D to help you determine the required battery capacity.<br />

The suction line carries the liquid from the sampling point to the<br />

<strong>Glacier</strong> pump tubing. The <strong>Glacier</strong> is designed to use:<br />

• 1 /4-inch (6-mm) I.D. Vinyl tubing<br />

• 3 /8-inch (9-mm) I.D. Vinyl tubing<br />

• 3 /8-inch (9-mm) I.D. PTFE tubing with a protective<br />

polyethylene jacket<br />

Note<br />

Selecting 1 /4-inch (6-mm) suction line disables the Bottle Full<br />

detection.<br />

Note<br />

The vinyl suction line contains a very low ppm (parts per million)<br />

level of phenols. If this affects your samples, use the<br />

PTFE suction line.<br />

The strainer reduces the possibility of debris plugging the<br />

suction line. Three types are available:<br />

• 3 /8-inch standard weighted polypropylene strainer<br />

2-9