Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />
Section 2 Installation<br />
CAUTION<br />
Never use a DC extension cable without first consulting with a<br />
Teledyne <strong>Isco</strong> Service Technician. They will advise you of the<br />
proper wire gauge for the length you require.<br />
The refrigerator compressor is equipped with a battery controller<br />
that will cut out to protect the compressor and the battery when<br />
the voltage drops to 10.4–11.3 VDC. The controller will cut in at<br />
11.7–12.5 VDC.<br />
CAUTION<br />
Never charge the battery while connected to the <strong>Glacier</strong>.<br />
Over-voltages could damage the electronics.<br />
Note<br />
The refrigerator housing has a High/Low power switch<br />
mounted near the DC input power connector. This High/Low<br />
switch is unused. The switch must remain in the Low position<br />
for normal operation.<br />
2.7.2 Battery<br />
Recommendations<br />
2.8 Connect a Suction<br />
Line and Strainer<br />
Before each sampling program, the battery should be exchanged<br />
with a fully-charged battery.<br />
Power consumption is mostly determined by the ambient temperature.<br />
Higher ambient temperatures will in turn create a higher<br />
demand for power. A battery selection guide is provided in<br />
Appendix D to help you determine the required battery capacity.<br />
The suction line carries the liquid from the sampling point to the<br />
<strong>Glacier</strong> pump tubing. The <strong>Glacier</strong> is designed to use:<br />
• 1 /4-inch (6-mm) I.D. Vinyl tubing<br />
• 3 /8-inch (9-mm) I.D. Vinyl tubing<br />
• 3 /8-inch (9-mm) I.D. PTFE tubing with a protective<br />
polyethylene jacket<br />
Note<br />
Selecting 1 /4-inch (6-mm) suction line disables the Bottle Full<br />
detection.<br />
Note<br />
The vinyl suction line contains a very low ppm (parts per million)<br />
level of phenols. If this affects your samples, use the<br />
PTFE suction line.<br />
The strainer reduces the possibility of debris plugging the<br />
suction line. Three types are available:<br />
• 3 /8-inch standard weighted polypropylene strainer<br />
2-9