Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Glacier Transportable Sampler - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />
Section 1 Introduction<br />
1.3 Technical<br />
Specifications<br />
Table 1-1 lists the technical specifications for the <strong>Glacier</strong> <strong>Transportable</strong><br />
<strong>Sampler</strong>.<br />
Table 1-1 <strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong> Specifications<br />
Size (H W D) 25 15 24 in. 63 38 60 cm<br />
(Does not include optional mobility kit)<br />
Weight 60 lb. 28 kg<br />
(Includes refrigerator, controller, empty bottle, and bottle retainer. Does not<br />
include battery or optional mobility kit)<br />
Power<br />
The mains line cord is the<br />
disconnect device.<br />
Installation Category<br />
Pollution Degree 4<br />
Maximum Altitude 1<br />
2000 meters<br />
Pump<br />
Intake Purge<br />
Tubing Life Indicator<br />
Intake Suction Line:<br />
Length<br />
Material<br />
Inside Dimension<br />
Pump Tubing Life<br />
Maximum Suction Lift<br />
Typical Repeatability<br />
Typical Line Velocity:<br />
3 ft (1 m)<br />
10 ft (3.1 m)<br />
15 ft (4.6 m)<br />
Liquid Presence Detector<br />
Controller<br />
DC Power Configuration: 12 VDC, 6 amperes<br />
Dual-power Configuration: 12 VDC, 6 amperes, or<br />
87 to 264 VAC, 47 to 63 Hz, 2 amperes<br />
II<br />
Before and after each sample collection<br />
Provides message to change tubing<br />
3 to 99 feet 1 to 30 m<br />
Vinyl or PTFE line<br />
0.375 in. or 0.250 in. 0,952 or 0,635 cm<br />
2,000 samples typical<br />
26 feet<br />
± 10 ml<br />
2.9 ft/sec 0,88 m/sec<br />
2.5 ft/sec 0,76 m/sec<br />
1.9 ft/sec 0,58 m/sec<br />
Non-wetted, non-conductive sensor for automatic detection of liquid<br />
Weight 8 pounds 3.6 kg<br />
Dimensions 10 x 12.5 x 10 inches 26 x 32 x 25 cm<br />
Operational Temperature 32° to 120° F 0° to 49° C<br />
Enclosure Rating NEMA 4x, 6 IP 67<br />
Program Memory<br />
Flow Meter Signal Requirements<br />
Real Time Clock Accuracy<br />
Refrigerator<br />
Non-volatile ROM<br />
5 to 15 volt DC pulse or 25 millisecond or longer isolated contact closure<br />
1 minute per month typical<br />
Sample maintenance temperature 37.4 ± 1.8° F 3.0 ± 1.0° C<br />
Software<br />
Sample Frequency Selection<br />
1 minute to 9,999 minutes in 1 minute increments 1 to 9,999 flow pulses<br />
1-5