<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 4 Maintenance and Special Features<br />

4.2 Needle Alignment When replacing the needle it is necessary to make sure that the<br />

needle aligns with the bottle to guarantee proper sealing and<br />

operation of the mechanism. Use the following procedure to<br />

ensure alignment.<br />

Figure 4-2 Needle Assembly Completed<br />

Figure 4-3 Needle Assembly Parts<br />

1. With the front shell still removed and the sample supply<br />

turned off, visually align the needle over the bottle rack,<br />

both side-to-side and front-to-back, leaving the gray nut at<br />

the needle support loose. Set the sampler to run a single<br />

bottle manual sample with no strainer purge.<br />

4-3