Section 4 Maintenance and Special Features<br />
4.2 Needle Alignment When replacing the needle it is necessary to make sure that the<br />
needle aligns with the bottle to guarantee proper sealing and<br />
operation of the mechanism. Use the following procedure to<br />
ensure alignment.<br />
Figure 4-2 Needle Assembly Completed<br />
Figure 4-3 Needle Assembly Parts<br />
1. With the front shell still removed and the sample supply<br />
turned off, visually align the needle over the bottle rack,<br />
both side-to-side and front-to-back, leaving the gray nut at<br />
the needle support loose. Set the sampler to run a single<br />
bottle manual sample with no strainer purge.<br />
4-3