Section 4 Maintenance and Special Features<br />

4.1 Replacing the<br />

Needle Assembly<br />

It is possible to replace the needle assembly in the field, if that<br />

should become necessary. Before attempting this, please<br />

read the following warning:<br />

WARNING<br />

Disconnect power completely before attempting needle<br />

replacement. The needle drive mechanism will cause<br />

serious injury to your hand if the sampler starts while you<br />

are working on it.<br />

1. Note that the tower housing consists of two pieces that fit<br />

together like a clam shell. Remove the front half of the<br />

tower housing. The two sections of the tower housing fit<br />

together inside an H-shaped molding. There are four<br />

thumbscrews that hold the front on. You should be able to<br />

see the needle assembly clearly after removing the housing.<br />

Figure 4-1 Tower Mechanism Showing Location of the Needle Assembly<br />

4-1