Section 3 Installation of the <strong>Sampler</strong><br />

an icon of a flow meter just beneath it. Connect the other end of<br />

the cable to the six-pin connector labeled SAMPLER on the flow<br />

meter.<br />

3.2.1 Connection to a<br />

Non-<strong>Isco</strong> Flow Meter<br />

You can connect certain non-<strong>Isco</strong> flow meters directly to a<br />

<strong>6100CR</strong> for flow-paced sampling. The flow meter must have an<br />

isolated contact closure of at least 25 milliseconds to provide<br />

acceptable flow pulses to the sampler. The frequency of the<br />

contact closure must be directly proportional to total flow.<br />

Connect the flow meter pulse output to the A and C pins of the<br />

Flow Meter connector on the <strong>6100CR</strong>. Teledyne <strong>Isco</strong> has an<br />

interface kit with a six-pin M/S connector that mates with the<br />

connector on the sampler. Wire connections from the non-<strong>Isco</strong><br />

flow meter to the six-pin connector and plug it into the sampler.<br />

Teledyne <strong>Isco</strong> also has a six-pin M/S connector wired to a 22 foot<br />

cable terminated in two wires. The black wire connects to pin A<br />

and the white wire connects to pin C. The following table shows<br />

the connections for the Flow Meter connector on the sampler.<br />

Pin Signal<br />

A<br />

+12 VDC<br />

B<br />

Common<br />

C<br />

Flow Pulses In<br />

D<br />

Bottle Number Out<br />

E<br />

Event Mark Out<br />

F<br />

Inhibit In<br />

Note that you will be unable to communicate anything other<br />

than the flow pulse contact closure with a non-<strong>Isco</strong> flow meter.<br />

The non-<strong>Isco</strong> flow meter will not be able to interpret event and<br />

bottle number information. If the flow pulse generated by the<br />

contact closure on the flow meter is not compatible with Teledyne<br />

<strong>Isco</strong>’s standard, contact Teledyne <strong>Isco</strong>’s Special Products<br />

Department for an appropriate interface device.<br />

Flow Meters with Non-Pulsed Flow Outputs - You can also<br />

use the <strong>6100CR</strong> with flow meters that have outputs other than a<br />

flow pulse. One common output type is the 4-20 mA current loop<br />

used for many types of industrial control equipment. To use the<br />

4-20 mA signal to pace the <strong>6100CR</strong>, you must use the 4-20 mA<br />

<strong>Sampler</strong> Input Interface. This device converts the constant<br />

4-20 mA current into flow pulses acceptable to the <strong>6100CR</strong>.<br />

If you use the <strong>6100CR</strong> with other manufacturers’ flow meters,<br />

please note that Teledyne <strong>Isco</strong> cannot assume any liability for<br />

operation or results obtained with the <strong>6100CR</strong> and other manufacturers’<br />

equipment.<br />