Section 3 Installation of the <strong>Sampler</strong><br />
an icon of a flow meter just beneath it. Connect the other end of<br />
the cable to the six-pin connector labeled SAMPLER on the flow<br />
meter.<br />
3.2.1 Connection to a<br />
Non-<strong>Isco</strong> Flow Meter<br />
You can connect certain non-<strong>Isco</strong> flow meters directly to a<br />
<strong>6100CR</strong> for flow-paced sampling. The flow meter must have an<br />
isolated contact closure of at least 25 milliseconds to provide<br />
acceptable flow pulses to the sampler. The frequency of the<br />
contact closure must be directly proportional to total flow.<br />
Connect the flow meter pulse output to the A and C pins of the<br />
Flow Meter connector on the <strong>6100CR</strong>. Teledyne <strong>Isco</strong> has an<br />
interface kit with a six-pin M/S connector that mates with the<br />
connector on the sampler. Wire connections from the non-<strong>Isco</strong><br />
flow meter to the six-pin connector and plug it into the sampler.<br />
Teledyne <strong>Isco</strong> also has a six-pin M/S connector wired to a 22 foot<br />
cable terminated in two wires. The black wire connects to pin A<br />
and the white wire connects to pin C. The following table shows<br />
the connections for the Flow Meter connector on the sampler.<br />
Pin Signal<br />
A<br />
+12 VDC<br />
B<br />
Common<br />
C<br />
Flow Pulses In<br />
D<br />
Bottle Number Out<br />
E<br />
Event Mark Out<br />
F<br />
Inhibit In<br />
Note that you will be unable to communicate anything other<br />
than the flow pulse contact closure with a non-<strong>Isco</strong> flow meter.<br />
The non-<strong>Isco</strong> flow meter will not be able to interpret event and<br />
bottle number information. If the flow pulse generated by the<br />
contact closure on the flow meter is not compatible with Teledyne<br />
<strong>Isco</strong>’s standard, contact Teledyne <strong>Isco</strong>’s Special Products<br />
Department for an appropriate interface device.<br />
Flow Meters with Non-Pulsed Flow Outputs - You can also<br />
use the <strong>6100CR</strong> with flow meters that have outputs other than a<br />
flow pulse. One common output type is the 4-20 mA current loop<br />
used for many types of industrial control equipment. To use the<br />
4-20 mA signal to pace the <strong>6100CR</strong>, you must use the 4-20 mA<br />
<strong>Sampler</strong> Input Interface. This device converts the constant<br />
4-20 mA current into flow pulses acceptable to the <strong>6100CR</strong>.<br />
If you use the <strong>6100CR</strong> with other manufacturers’ flow meters,<br />
please note that Teledyne <strong>Isco</strong> cannot assume any liability for<br />
operation or results obtained with the <strong>6100CR</strong> and other manufacturers’<br />
equipment.<br />