Section 3 Installation of the <strong>Sampler</strong><br />
Flow<br />
Arrow<br />
Figure 3-3 Filter, Disassembled<br />
Filter Screen Element<br />
Arrow<br />
3.1.2 Preparation of Tubing<br />
and Connectors<br />
The tubing used for the sample delivery line is terminated with<br />
specially-designed connectors in an optimized system that is both<br />
leak-tight and easy to install.<br />
Figure 3-4 Fitting, Disassembled<br />
Note<br />
Normally there is no need to disassemble the tubing fittings. In<br />
fact, it is better to avoid disassembly, as it increases the risk of<br />
loss of small internal parts, or of incorrect reassembly. In actual<br />
practice, it is only necessary to groove the tubing with the scoring<br />
tool, insert the tubing into the fitting after loosening the blue<br />
end, and then re-tighten to secure the tubing. Make sure the<br />
tubing bottoms out in the fitting. Tighten the fitting nut only to<br />
the point where the nut just stops at the ridge of the fitting<br />
body. Do not overtighten; the fittings are Teflon ® and the<br />
threads will be destroyed!<br />
3-4