6100CR Refrigerated Sampler - Isco
6100CR Refrigerated Sampler - Isco
6100CR Refrigerated Sampler - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />
Section 3 Installation of the <strong>Sampler</strong><br />
The sample line runs to the <strong>6100CR</strong> through a filter.<br />
Incoming Sample<br />
Filter Bypass<br />
To Drain<br />
Figure 3-1 Location of Sample Line Filter on Refrigerator<br />
Sample Overflow<br />
to Drain<br />
Incoming Line Filter To Pressure Regulator<br />
Figures 3-2 and 3-3 provide a close-up of the installed filter. Note<br />
that the filter must be installed in the proper direction. The<br />
knurled endcaps have flow direction arrows stamped on them,<br />
and this is how you can tell if the filter has been installed in the<br />
proper direction. Refer to section 3.1.2 for tubing preparation.<br />
3-2