<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 2 Programming<br />

Or:<br />

UNIT OFF AT<br />

HH:MM DDMMMYY<br />

4. When you activate a sampling program by pressing the<br />

Run key, the following display will appear on the log:<br />

PROGRAM ACTIVATED<br />

HH:MM DDMMMYY<br />

When the sampling sequence actually begins, the following<br />

display will appear on the log:<br />

PROGRAM STARTED AT<br />

HH:MM DDMMMYY<br />

The following log entry will display for a sampling event:<br />

SAMPLE: 1 HH:MM<br />

BOTTLE: 1 DD MMM<br />

5. If you press the Right Arrow key again, the following log<br />

entry will appear:<br />

SOURCE: FLOW<br />

The possible Sources are: TIME, FLOW, START, MULTI-<br />

PLE, and MANUAL.<br />

If there was a problem and the unit was unable to take a<br />

sample at that time, this message would appear instead:<br />

SOURCE: FLOW<br />

ERROR: RACK JAM (example)<br />

The possible errors are: RACK JAM, TOWER JAM, RACK<br />

FULL, UNIT OFF, VALVE JAM, DISABLED, and POWER<br />

FAIL.<br />

When the sampling sequence ends, the log will display:<br />

PROGRAM FINISHED AT<br />

HH:MM DDMMMYY<br />

When the sampling program has finished (or was stopped)<br />

and the <strong>6100CR</strong> has returned to the standby state, the log<br />

2-18