<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />
Section 2 Programming<br />
select YES. See Step 8.<br />
ENTER START TIME<br />
YES NO<br />
8. START TIME is the time the sampling routine starts.<br />
Enter a time later than the present time, or you will be<br />
asked to reenter the Start Time. In Time mode, The<br />
<strong>6100CR</strong> will take the first sample at Start Time. In Flow<br />
mode, the <strong>6100CR</strong> will only take a sample at Start Time if<br />
you enabled that option in the Programming sequence with<br />
Flow Pacing selected. If not, the <strong>6100CR</strong> will simply begin<br />
counting down the flow pulses before the first sample.<br />
START TIME<br />
HH:MM MMM/DD<br />
Note<br />
When you finish programming the <strong>6100CR</strong>, you must press<br />
RUN for the sampler to run its program. Otherwise, the program<br />
will never run, even after the start time passes.<br />
2.5 The RUN Key 1. If you want to start at the beginning of the rack (bottle 1),<br />
press Rack Reset before pressing Run. After you press<br />
Run, the following display will appear:<br />
START AT BOTTLE<br />
X<br />
The value of X will be the next available bottle. If the bottle<br />
rack is in the “home” position, this will be bottle 1. If the<br />
rack is on the last bottle (number 24), the sampler will<br />
reset the rack to the “home” position and the display will<br />
show the bottle “1.”<br />
Pressing On/Off will turn the sampler off. Pressing<br />
Clear/Exit will return the sampler to the standby state;<br />
the Run key must be pressed again to operate the sampler.<br />
You can press Enter to select the displayed bottle to start<br />
the sampling routine, or you may enter a higher bottle<br />
number. The highest number you can enter is 24. You cannot<br />
enter a smaller number than that displayed.<br />
The number entered is the maximum number of sample<br />
events that the sampler will perform. If the sampler starts<br />
at a bottle other than bottle #1, the sample program may<br />
2-11