Section 2 Programming<br />
should always discharge into a normal, non-pressurized<br />
open drain.<br />
2. If you select YES, the following will appear:<br />
PUT DISCHARGE TUBE IN DRAIN<br />
Alternating with:<br />
Make sure that the sampler drain line is in a suitable<br />
drain and press the Manual Purge key. This will start liquid<br />
flow through the sample cooler and discharge through<br />
the drain line.<br />
Followed by:<br />
THEN,<br />
PRESS MANUAL PURGE<br />
Observe the discharge line. After the air is purged from the<br />
sample cooler, you should only observe a few bubbles in the<br />
sample line. When the air is purged, push Clear/Exit, and<br />
the flow will stop.<br />
Followed by:<br />
WHEN NO BUBBLES<br />
PRESS CLEAR/EXIT<br />
PUT DISCHARGE TUBE<br />
IN GRAD CYLINDER<br />
Alternating with:<br />
THEN,<br />
PRESS MANUAL SAMPLE<br />
Place the discharge tube into a beaker or graduated cylinder.<br />
Place the beaker at the same level as the discharge<br />
tube will be in normal operation. Then press the Manual<br />
Sample key to start the 50 second flow for calibration. The<br />
display will show:<br />
COLLECTING 100 ML<br />
TEST VOLUME XX<br />
2-3