Section 1 Introduction<br />

2. Unpack the box carefully, checking for any possible shipping<br />

damage.<br />

3. Remove the tower shipping support. Reinstall the black<br />

knob taken off when removing the support.<br />

1.3.1 Under the Controller<br />

Cover<br />

When you have opened the cover over the controller, the most<br />

prominent features inside are the keypad and display. You<br />

program the sampler using the keypad and prompts from the<br />

display.<br />

Inside the refrigerator behind the sample bottle rack is a rectangular<br />

protrusion with a yellow warning label on it. This is the<br />

housing for the fill needle. When the sampler is in operation, the<br />

fill needle raises and lowers automatically to inject the liquid<br />

sample into the bottles in the rack.<br />

The warning label is to warn you against putting your hand<br />

between the tower and the bottle rack. You can be seriously<br />

injured by the needle coming down when the sampler is in operation.<br />

WARNING<br />

Never place your hand or any other body parts between the<br />

needle assembly and the bottle rack. If the sampler starts<br />

unexpectedly, it will lower the needle to fill a bottle. You<br />

could suffer serious personal injury if your hand is caught<br />

between the needle and the bottle rack. The needle comes<br />

down with enough force to cause serious injury. In order to<br />

avoid damage, always make sure bottle rack is in place,<br />

bottles are in all 24 positions, and the shut-off handles on<br />

all bottles are horizontal.<br />

1.3.2 The Bottle Rack Teledyne <strong>Isco</strong> ships the bottle rack separately from the sampler<br />

for protection. Because of the carrier’s insulating properties you<br />

can pack the bottle rack with ice (or gel packs if you are<br />

shipping) to keep the samples cool from when they are gathered<br />

to when they are analyzed at the laboratory. (Gel packs are a<br />

packaged cooling agent.)<br />

It is very important to keep the samples cool to keep them representative;<br />

heat drives volatile organic compounds out of solution.<br />

The metal plate holding the bottles inside the bottle rack keeps<br />

the bottles in place during shipment and use. It prolongs the<br />

melting time of the ice. The knurled knob with the threaded<br />

shaft attaches the bottle rack to the controller. Never operate the<br />

sampler without the stainless steel cover in place and the knob<br />

securely tightened. Before installing the bottle rack, look at the<br />

tops of the bottles. Make sure all the valve handles on the bottles<br />

are turned so they are perpendicular to the standing bottles, or<br />

facing 3 and 9 o’clock. (See Figure 1-2.)<br />

This is important, not only for proper alignment for sampling,<br />

but also to ensure thst all bottles are closed and free from contamination.<br />

During normal operation, the sampler may occasionally<br />

leak a small amount of sample around the O-ring on the<br />

1-4