D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

3.5.3 STORE The store key is operational when in the programmed gradient<br />

mode. It is used to save the program gradient parameters and<br />

exit the programming (parameter entry) mode.<br />

3.5.4 LIMITS The controller allows the user to set the minimum and maximum<br />

flow rate limits, the minimum and maximum pressure limits,<br />

and the maximum rate the pump will run while controlling the<br />

pressure in constant pressure mode.<br />

• When using a single controller to operate multiple<br />

pumps, you need to select the appropriate pump before<br />

setting any pump parameters. The available pumps will<br />

be displayed above the softkeys. These selections correspond<br />

with the connector that the pump control cable is<br />

plugged into on the rear panel of the pump controller. To<br />

select a pump, press the softkey under the pump designation.<br />

The top line of the screen will indicate the pump<br />

which is currently selected.<br />

• The high and low limits you set cannot exceed the pump<br />

specifications.<br />

To set the limits<br />

1. Press the LIMITS key on the keypad. To display the Limits<br />

menu, Figure 3-7, enter the number of the limit you wish<br />

to program using the number keys. This will cause one of<br />

five limit setpoint menus to appear, MAX PRESS, MIN<br />

PRESS, MAX FLOW, MIN FLOW, or FLOW LIMIT. The<br />

MAX PRESS limit setpoint menu is shown in Figure 3-8.<br />

Figure 3-7 Limits menu<br />

LIMITS:<br />

CURRENT PUMP<br />

1. MAX PRESS _____<br />

2. MIN PRESS______<br />

PUMP A<br />

3. MAX FLOW____<br />

4. MIN FLOW_____<br />

5. FLOW LIMIT_____<br />

PREVIOUS<br />

A B C D<br />

MAX PRESS_____PSI<br />

1. SET PRESS VALUE<br />

Figure 3-8 Limits setpoint (Max Press) menu<br />

ON ON OFF<br />

DISPLAY* ALARM* SHUTDOWN* PREVIOUS*<br />

A B C D<br />

Note<br />

These features are discussed later in this section under Limits<br />

Programming Options.<br />

2. Press the number 1 key to set the value. A message will<br />

blink on the right side of the screen, prompting you to<br />

enter the selected limit.<br />

3. Enter the desired limit setpoint, using the number keys.<br />

4. Press the ENTER key to save the value.<br />

3-14