Section 3 Single & Multiple Independent Pump Operation<br />

DIFF. MODES – The standard pressure transducer MUST be<br />

connected to pump for correct operation. This feature allows the<br />

use of other analog inputs for pressure input to the pump.<br />

Press 1, OFF, to turn off this feature.<br />

Press 2, 0 to 50 ANLG1, to use a 5 volt 50 psi<br />

transducer on analog input ANLG1.<br />

Press 3, Custom ANLG1, to use 5 volt custom pressure<br />

transducer on analog input ANLG1.<br />

Press 4, 500 ANLG2, to use a 5 volt 500 psi pressure<br />

transducer on analog input ANLG2.<br />

Press 5, 5000 ANLG3, to use a 5 volt 5000 psi<br />

transducer on analog input ANLG3.<br />

PRESS. INTEGRAL – This feature lets the user turn on/off the<br />

integral function for the pump control.<br />

3.5 Front Panel Keys In addition to the menu options, certain features are conveniently<br />

located on the front panel keyboard. These include:<br />

• HOLD, 3.5.1<br />

• RECALL, 3.5.2<br />

• STORE, 3.5.3<br />

• LIMITS, 3.5.4<br />

• RAPID PRESS (Rapid Pressure), 3.5.5<br />

• HELP, 3.5.6<br />

• ACCESS CTRL (Accessories Control), 3.5.7<br />

• ZERO PRESS (Zero Pressure), 3.5.8<br />

HOLD, RECALL, and STORE are all used when in program gradient<br />

mode.<br />

3.5.1 HOLD The HOLD key is used while a gradient is running. When the<br />

HOLD key is pressed, this will freeze the program clock and<br />

maintain present gradient parameters.<br />

To continue the gradient, press the HOLD or RUN key.<br />

3.5.2 RECALL The RECALL key is used to recall a previously programmed gradient.<br />

When you select this option, the controller automatically<br />

loads the gradient and switches the pump to program gradient<br />

mode. The RECALL key can be used when the pump(s) are<br />

stopped or when in the HOLD mode.<br />

1. Press the RECALL key and use the number keys to enter<br />

the number name of the gradient you wish to recall. Press<br />

the Enter key.<br />

2. If you enter a number of a gradient which does not exist,<br />

the controller briefly displays the message “FILE NUM-<br />

BER DOES NOT EXIST.” It then assumes you will be creating<br />

a new gradient under that number and displays the<br />

program gradient run screen.<br />

3-13