Section 3 Single & Multiple Independent Pump Operation<br />

3. Press number 6, VALVE. The number for the selected valve<br />

type will be blinking.<br />

4. Press the number 1, 2 or 3 key to select a desired option<br />

and the number 1, 2, 3 will blink indicating which valve<br />

type is selected.<br />

5. To exit this menu, press the softkey D, PREVIOUS, to<br />

return to the main menu.<br />

3.4 Second Menu The Second Menu gives you access to various functions. Access<br />

the Second Menu by pressing softkey A, MORE, two times.<br />

1. POOR FILL<br />

2. DIAGNOSTIC MENU<br />

3. PRESS. CALIBRATION<br />

4. DIFF. MODES<br />

5. PRESS. INTEGRAL<br />

Figure 3-6 Second Menu<br />

A B C D<br />

3.4.1 Second Menu Menu<br />

Options<br />

POOR FILL – In the constant pressure mode, this feature<br />

allows you to set a fill point as a percentage of pump volume. If<br />

this volume percentage is not reached after a refill and re-pressurization,<br />

the system sounds an alarm and stops the pump.<br />

DIAGNOSTIC MENU – Displays an additional menu containing<br />

testing options<br />

LCD TEST<br />

This feature will cycle the display. Press any key to<br />

stop test.<br />

ANALOG INPUT<br />

This feature will display all the analog input signals on<br />

the display.<br />

MEMORY TEST – This feature will test the memory<br />

for errors.<br />

KEYPAD TEST – This feature will test the keypad.<br />

Press “EXIT” to exit.<br />

MOTOR CONTROL/LIMITS – This feature will test the<br />

digital position control system. Press 1, 2, 3 for the<br />

pump wanted for test. Press “UP 100” to move the<br />

pump’s cylinder up 100 counts. Press “DOWN 100” to<br />

move the pump’s cylinder down 100 counts. Press<br />

“EXIT” to exit this function.<br />

SERIAL TEST – This feature will test the serial<br />

channel. Once this feature is active the software will<br />

serial send “*******SERIAL TEST XXXX*********”<br />

through the serial port. The “REC.” message will show<br />

any serial input characters on the display.<br />

PRESS. CALIBRATION – This feature is used for calibration<br />

only. If an incorrect number is entered with this feature, the<br />

pumps will NOT display/run at the correct pressure. Press<br />

“CalA” (“Cal”, “CalC”) and enter the calibration number for the<br />

needed pump. Press “Prev” to exit this feature.<br />