Section 3 Single & Multiple Independent Pump Operation<br />
To designate independent<br />
control of multiple pumps<br />
1. Press the orange MENU key.<br />
2. Press softkey A, MORE.<br />
3. Press number 4, MULTI PUMP. The multi-pump menu<br />
will then be displayed.<br />
4. Press number 4, INDEPENDENT. The controller will set<br />
the pumps to Independent mode. Number 4 will blink,<br />
indicating that INDEPENDENT mode is selected.<br />
5. Select the HOLD PRESS or NORMAL mode of operation.<br />
Press softkey A to toggle between the two modes.<br />
HOLD PRESS: After the pump is empty in constant<br />
pressure mode, if the outlet pressure rises past the set<br />
point the pump will restart and run the system to the set<br />
point pressure.<br />
NORMAL: This feature shuts the pump off if the pumps<br />
runs empty in constant pressure mode.<br />
Once the pumps have been set to this mode, they will<br />
operate independently from one another. Each pump will<br />
operate at its defined limit and rate. One pump may be<br />
operating in constant flow, the other in constant pressure.<br />
Independent mode is the default setting for the pump.<br />
When a command such as stop or refill is pressed, a menu<br />
will appear asking you to designate the pump to stop or<br />
refill. Only the designated pump will stop, the other pumps<br />
will continue in the pumping application you have them set<br />
for.<br />
6. Return to the run screen by pressing softkey D three times.<br />
Then press softkey D (select pump) to display the select<br />
pump screen. This screen will show each pump’s pertinent<br />
information and allow you to select any pump for programming<br />
changes.<br />
7. To select B pump from the select pump screen, press softkey<br />
B. This will display the B pump run screen. From here,<br />
you can start pump B or change the operating parameters.<br />
3.3.10 Total Volume Reset This feature will reset the total volume display used in the continuous<br />
flow, or modifier option to zero.<br />
To reset total volume to zero<br />
1. Press the orange MENU key.<br />
2. Press softkey A, MORE.<br />
3. Press number 5, TOTAL VOL RESET.<br />
The volume will then be reset to zero.<br />
3.3.11 Valve This feature will designate the type of valves (passive, active or<br />
electric) being used for the flow operation. (If “active value” or<br />
“electric” is selected, the controller will match the pressure more<br />
closely before switching delivery pumps.)<br />
To select valve type<br />
1. Press the orange MENU key.<br />
2. Press softkey A, MORE.<br />
3-11