D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />
Section 3 Single & Multiple Independent Pump Operation<br />
Table 3-1 Key functions in the Multi-pump Mode<br />
Key Display Option Description<br />
A NORMAL Uses a finer (slower) pressure<br />
match control when<br />
switching from one pump to<br />
the other.<br />
FAST<br />
Uses a coarser (faster)<br />
pressure match control<br />
when switching from one<br />
pump to the other.<br />
B NORMAL PRESS Uses pressure matching<br />
when switching from one<br />
pump to the other<br />
LOW PRESS<br />
Uses no pressure matching<br />
when switching from one<br />
pump to the other.<br />
C DELIVER Sets the pump into the<br />
delivery mode of operation.<br />
RECEIVE<br />
Sets the pump into the<br />
receive mode of operation.<br />
6 MIN/MAX POINTS Sets the fill and refill marks<br />
that are used with both continuous<br />
flow modes.<br />
Continuous Flow (constant<br />
pressure mode)<br />
Modifier (constant pressure<br />
mode)<br />
Independent Control of up to<br />
Three <strong>Pumps</strong><br />
A D <strong>Series</strong> continuous flow pumping system in constant pressure<br />
mode, will consist of two D <strong>Series</strong> syringe pumps, and a valve<br />
accessory package, all regulated by one controller. To connect two<br />
D <strong>Series</strong> pumps you will need to use a continuous flow check<br />
valve package (P/N 68-1247-059) or a continuous flow air driven<br />
valve package (P/N 68-1247-058 for 100D/260D, 68-1247-061 for<br />
500D, or 68-1247-104 for 1000D). Consult with the factory when<br />
configuring 65D pumps for continuous flow. Installation and<br />
operating instructions for this system are located in Section 8.<br />
In this mode the softkeys toggle between the options described in<br />
Table 3-1.<br />
A pump modifier system includes two D <strong>Series</strong> syringe pumps,<br />
regulated by one controller and outlet check valves to prevent<br />
system back flow into the pump. To connect the 100D/260D<br />
pumps, you will need a modifier addition kit (P/N 68-1247-079).<br />
Consult with the factory when configuring 65D pumps as a modifier<br />
system. Installation and operating instructions for modifier<br />
systems are located in Section 9.<br />
In this mode the softkeys toggle between the options described in<br />
Table 3-1.<br />
A D <strong>Series</strong> syringe pump controller can run three syringe pumps<br />
independently of each other in either constant pressure or constant<br />
flow mode or any combination of the two. To set up this<br />
option use the procedure detailed below.<br />
3-10