Section 3 Single & Multiple Independent Pump Operation<br />

Table 3-1 Key functions in the Multi-pump Mode<br />

Key Display Option Description<br />

A NORMAL Uses a finer (slower) pressure<br />

match control when<br />

switching from one pump to<br />

the other.<br />

FAST<br />

Uses a coarser (faster)<br />

pressure match control<br />

when switching from one<br />

pump to the other.<br />

B NORMAL PRESS Uses pressure matching<br />

when switching from one<br />

pump to the other<br />

LOW PRESS<br />

Uses no pressure matching<br />

when switching from one<br />

pump to the other.<br />

C DELIVER Sets the pump into the<br />

delivery mode of operation.<br />

RECEIVE<br />

Sets the pump into the<br />

receive mode of operation.<br />

6 MIN/MAX POINTS Sets the fill and refill marks<br />

that are used with both continuous<br />

flow modes.<br />

Continuous Flow (constant<br />

pressure mode)<br />

Modifier (constant pressure<br />

mode)<br />

Independent Control of up to<br />

Three <strong>Pumps</strong><br />

A D <strong>Series</strong> continuous flow pumping system in constant pressure<br />

mode, will consist of two D <strong>Series</strong> syringe pumps, and a valve<br />

accessory package, all regulated by one controller. To connect two<br />

D <strong>Series</strong> pumps you will need to use a continuous flow check<br />

valve package (P/N 68-1247-059) or a continuous flow air driven<br />

valve package (P/N 68-1247-058 for 100D/260D, 68-1247-061 for<br />

500D, or 68-1247-104 for 1000D). Consult with the factory when<br />

configuring 65D pumps for continuous flow. Installation and<br />

operating instructions for this system are located in Section 8.<br />

In this mode the softkeys toggle between the options described in<br />

Table 3-1.<br />

A pump modifier system includes two D <strong>Series</strong> syringe pumps,<br />

regulated by one controller and outlet check valves to prevent<br />

system back flow into the pump. To connect the 100D/260D<br />

pumps, you will need a modifier addition kit (P/N 68-1247-079).<br />

Consult with the factory when configuring 65D pumps as a modifier<br />

system. Installation and operating instructions for modifier<br />

systems are located in Section 9.<br />

In this mode the softkeys toggle between the options described in<br />

Table 3-1.<br />

A D <strong>Series</strong> syringe pump controller can run three syringe pumps<br />

independently of each other in either constant pressure or constant<br />

flow mode or any combination of the two. To set up this<br />

option use the procedure detailed below.<br />