D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

Pressure Example<br />

The maximum pressure desired is 510.2 ATM, (maximum<br />

allowed by the MODEL 260D) with a scale factor of 2.0 volts per<br />

100 ATM. The full-scale input voltage would be:<br />

510.2 ATM <br />

2.00 V<br />

= 10.204 volts<br />

100 ATM<br />

and MAX PRESS should be set to 510.2 ATM.<br />

Maximum Flow Rate<br />

Example<br />

The maximum flow rate desired is 25.0 ml/min with a scale factor<br />

of 5.0 volts per 20.0 ml/min. The full scale input voltage would be:<br />

5.0 V<br />

25.0 ml/min <br />

= 6.250 volts<br />

20.0 ml/min<br />

and MAX FLOW should be set to 25.0 ml/min.<br />

Wire Connections<br />

Two wires are required for analog control. The analog common or<br />

ground wire should be connected to the GND terminal under<br />

ANALOG INPUT of the ACCESSORY connector on the controller<br />

rear panel. The analog control or input wire should be connected<br />

to terminal 1 under ANALOG INPUT.<br />

If two pumps are used with the controller, the second analog<br />

control or input wire should be connected to terminal 2 under<br />

ANALOG INPUT. If three pumps are used with the controller,<br />

the third analog control or input wire should be connected to terminal<br />

3 under ANALOG INPUT or to auxiliary (P11) pin 15.<br />

When using one of the multi-pump operation modes, only the<br />

ANALOG INPUT terminal 1 needs to be connected.<br />

3.3.9 Multi-Pump When using multiple pumps, there are four multi-pump operating<br />

modes of delivery and an independent mode, including:<br />

• Continuous flow in constant flow mode.<br />

• Continuous flow in constant pressure mode.<br />

• Modifier addition in constant pressure mode.<br />

• Modifier addition in continuous flow, constant pressure<br />

mode.<br />

• Independent mode.<br />

Continuous Flow (constant<br />

flow mode)<br />

A D <strong>Series</strong> continuous flow pumping system in constant flow<br />

mode, will consist of two syringe pumps and a valve accessory<br />

package, all regulated by one controller.<br />

To connect two D <strong>Series</strong> pumps for a continuous flow pumping<br />

system, you will need a continuous flow check valve package (P/N<br />

68-1247-059) or a continuous flow air driven valve package (P/N<br />

68-1247-129 for 65D, 68-1247-058 for 100D/260D, 68-1247-061<br />

for 500D, or 68-1247-104 for 1000D). Installation and operating<br />

instructions for this system are located in Section 8.<br />

In this mode the softkeys toggle between the options described in<br />

Table 3-1.<br />

3-9