D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

Note<br />

This screen is momentarily displayed automatically each time<br />

the pump controller is turned from STANDBY to ON.<br />

To display the status<br />

3.3.8 External Analog<br />

Control<br />

Setting up external control<br />

1. Press the MENU key and then press softkey A, MORE.<br />

2. Press 2 to display the pump status.<br />

3. The first line displays the revision of the software.<br />

4. Lines 2, 3, and 4 display the type of pump connected to the<br />

A, B, and C pump connectors, respectively.<br />

The D <strong>Series</strong> syringe pump can be controlled externally by<br />

analog voltage in either constant flow or constant pressure mode.<br />

The input range is 0 – 11.5 volts (for all pumps), with a resolution<br />

of 5000 increments per volt.<br />

1. Select the desired operating mode (CONST FLOW or<br />

CONST control: PRESS) by pressing the appropriate key<br />

on the controller front panel.<br />

The mode you have selected, along with the current pump<br />

will be displayed in the upper left-hand corner of the<br />

screen. For example, if you have selected CONST PRESS,<br />

CP will be displayed, followed by a lowercase a, b, or c<br />

depending on the selected pump.<br />

To set parameters for a different pump, press softkey D,<br />

SELECT PUMP, and choose A for pump A, B for pump B,<br />

etc. Once you have changed pumps, the display will change<br />

back to the run screen, automatically.<br />

If you do not wish to select a different pump, press softkey<br />

D to return to the run screen.<br />

2. Press MENU, then softkey A for MORE options.<br />

3. Press number key 3 to select EXTERNAL control.<br />

4. Softkey B will toggle the EXTERNAL mode ON or OFF.<br />

5. If you wish to change the full-scale input voltage, press softkey<br />

A to SET RANGE. Then enter the new voltage range<br />

between 1.000 V and 11.500 V, and press ENTER to store<br />

the value. This voltage will correspond to the maximum<br />

flow rate or pressure value programmed by the LIMITS key.<br />

DV FS<br />

ML = input<br />

where:<br />

DV = desired value of pressure or flow rate.<br />

FS = full-scale input range.<br />

ML = max limit of pressure or flow rate<br />

input = external input voltage.<br />

3-8