D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Warnings, Cautions, and Notices<br />

- EXPLOSION WARNING -<br />

WARNING<br />

Teledyne <strong>Isco</strong> D <strong>Series</strong> <strong>Pumps</strong>, SFX 2-10 and<br />

SFX 220 Extractors are NOT EXPLOSION PROOF.<br />

Teledyne <strong>Isco</strong> SFX System and D <strong>Series</strong> <strong>Syringe</strong><br />

Pump Safety Note when using a flammable fluid<br />

The Teledyne <strong>Isco</strong> SFX system and syringe pumps must be placed within a properly<br />

operating vent hood (fume cupboard), when using ethane or any other flammable<br />

gas. Ensure that all SFE tubing connections are completely free of any gas<br />

leaks by performing the leak test using CO 2 (detailed in Section 2 of the D <strong>Series</strong><br />

pump manual, Section 5 of the SFX 2-10 manual and Section 6 of the SFX 2-10,<br />

220, and SFX 3560 manual). There must absolutely be NO gas leaks present before<br />

introducing the flammable gas. In a temperature-stable, leak-free system, the flow<br />

rate, as registered by the pump, should settle to a value below 0.01 ml/min after 15,<br />

minutes during a static extraction.<br />

Important: When using a Teledyne <strong>Isco</strong> extractor (either SFX 2-10, SFX<br />

220, SFX 3560) be absolutely sure the built-in venting fan (a brushless<br />

motor) is operating properly.<br />

The Teledyne <strong>Isco</strong> D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong> and the SFX units, use brush-type drive<br />

motors. Minor modifications to the pumps may render them safer, especially in the<br />

rare event of catastrophic piston seal failure. However, these modifications<br />

will not make these pumps explosion proof.<br />

•Remove the front and back cylinder covers located on the ball screw tower.<br />

This will allow any escaped gas to quickly dissipate away from the pump cylinder<br />

area and to reduce the amount entering the motor compartment.<br />

•Seal the syringe pump motor compartment with tape and purge it with a continuous<br />

flow of nitrogen (N 2 ) gas. This will also reduce the possibility of accumulating<br />

an explosive mixture around the motor and relays.<br />

•For further information, telephone Teledyne <strong>Isco</strong> at (800)775-2965, or fax<br />

(402)465-3085 to consult the Teledyne <strong>Isco</strong> Service Department.<br />

USE THE Teledyne <strong>Isco</strong> SFE SYSTEM AND SYRINGE PUMPS IN THESE<br />

POTENTIALLY HAZARDOUS APPLICATIONS AT YOUR OWN RISK!<br />