07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

3.3.2 Refill The refill option allows you to set the refill rate or have the pump<br />

automatically refill when a certain volume is reached.<br />

To set auto refill volume<br />

1. Press the orange MENU key.<br />

2. Press number 3, REFILL. The auto refill menu, Figure 3-3<br />

will be displayed.<br />

1. REFILL MARK<br />

2. AUTO REFILL<br />

3. REFILL RATE<br />

PUMP A<br />

VOLUME A:<br />

PUMP A:<br />

PUMP A:<br />

000.00mL<br />

OFF<br />

000.00mL/MIN<br />

PREVIOUS<br />

A B C D<br />

Figure 3-3 Refill menu<br />

Note<br />

If more than one pump is connected to the controller, pump B<br />

and pump C will be displayed on the fourth line. To select a<br />

pump, press the softkey under the pump designation.<br />

3. Press the number 1 key to set the volume for pump A.<br />

4. The units to the right of the symbol will blink, indicating<br />

that you should enter a volume. Use the number keys to<br />

enter an appropriate value and then press the ENTER key.<br />

Disable/Enable Auto Refill<br />

To set refill rate<br />

To exit<br />

The second line will display “OFF” or “ON”, indicating whether<br />

or not this feature is enabled for pump A (or the currently<br />

selected pump). Press the number 2 key to toggle this feature off<br />

and on for each pump, as desired.<br />

If auto refill is ON, then the pump will automatically switch to<br />

refill mode when the volume reaches the auto refill mark. After<br />

refilling, pumping will resume in the programmed mode. The<br />

ACCESSORY outputs, which drive powered valves, will switch in<br />

sequence.<br />

1. Press the number 3 key to set the refill rate for the designated<br />

pump. The refill rate can also be changed from the<br />

main screen while the pump is refilling.<br />

2. A message will blink on the screen prompting you to enter<br />

the selected refill rate.<br />

3. Enter the desired rate using the number keys.<br />

4. Press the ENTER key to save the value.<br />

1. To exit the refill menu, press softkey D, PREVIOUS. You<br />

will be returned to the main menu; and your selections will<br />

be saved.<br />

3.3.3 Power Failure [STOP] This feature allows you to dictate the activity of the pump in the<br />

event of a power failure. Use the number 4 key to toggle this<br />

feature to [STOP] or [CONT] to automatically resume pumping<br />

after power is restored.<br />

3-5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!