D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

The following options are available via the main menu, Figure<br />

3-1. Each option is discussed in the separate subsection indicated.<br />

• Units: Section 3.3.1.<br />

• Refill: Section 3.3.2.<br />

• Power failure: Section 3.3.3.<br />

• System reset: Section 3.3.4.<br />

• Display contrast: Section 3.3.5.<br />

• Serial options: Section 3.3.6.<br />

• Status Section: 3.3.7.<br />

• External analog control: Section 3.3.8.<br />

• Multi-pump mode: Section 3.3.9.<br />

• Total volume reset: Section 3.3.10.<br />

• Valve: Section 3.3.11.<br />

3.3.1 Setting Flow and/or<br />

Pressure Units<br />

For your convenience, the controller allows the user to set the<br />

displayed units on the pump. See Figure 3-2.<br />

PRESSURE UNITS=___________FLOW UNITS=__________<br />

1. ATM<br />

3. PSI<br />

5. mL/MIN 7. µL/MIN<br />

2. BAR<br />

4. kPA<br />

6. mL/HR 8. µL/HR<br />

PREVIOUS<br />

A B C D<br />

Figure 3-2 Units menu<br />

Setting the units<br />

1. Press the orange MENU key.<br />

2. Press number 1, UNITS.<br />

3. The units menu, Figure 3-2, will be displayed.<br />

4. Use numbers 1-4 to select the pressure units (ATM, BAR,<br />

PSI, kPa). Pressing the number associated with the unit<br />

will cause it to be displayed on the first line of the screen<br />

after PRESSURE UNITS=.<br />

5. Use numbers 5-8 (mL/MIN, mL/HR, µL/MIN, µL/HR) to set<br />

the flow rate units. Pressing the number associated with<br />

the unit will cause it to be displayed on the first line of the<br />

screen after FLOW UNITS=.<br />

6. To exit the UNITS menu, press softkey D, PREVIOUS.<br />

This will return you to the main menu screen. Your settings<br />

will be saved automatically for all pumps.<br />