D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong>: Model 65D<br />

Section 2D 65D Liquid System Connections & Accessories<br />

Reassembling the pump<br />

1. Place the cylinder/cooling jacket assembly over the piston<br />

and push rod assembly.<br />

2. Screw the assembly into the cylinder housing until the cylinder<br />

no longer turns. This indicates that it is snug against<br />

the piston.<br />

3. Unscrew the cylinder a minimum of 1 2 turn.<br />

4. Line the inlet and outlet cylinder cap ports up as you had<br />

them before. Turn the cooling jacket ports to the desired<br />

location and lock the cooling jacket to the cylinder by tightening<br />

the cooling jacket lock set screw.<br />

5. Lock the cylinder by tightening the cylinder locking screw.<br />

6. Replace the front cover and adjust both covers so they are<br />

flush with the cylinder housing.<br />

7. Reinstall tubing.<br />

8. Reconnect the pump pressure transducer cable.<br />

2D.4 Nitrogen Purge A thin film of liquid will wet the inside of the cylinder each time<br />

the piston travels up the cylinder. Corrosive liquids drying in the<br />

air can cause the cylinder to corrode. This corrosion occurs at<br />

varying rates based on the liquids and pumping conditions being<br />

employed. The lowest flow rate engenders the most corrosive<br />

pumping environment, as the deposited film has the longest time<br />

to corrode the inside of the cylinder wall.<br />

The 65D pump is equipped with a purge connector (Figure 2D-5).<br />

The purge connector enables the pump cylinder beneath the<br />

piston to be purged with nitrogen which inhibits cylinder corrosion<br />

and may increase the useful cylinder life. Figure 2D-5<br />

shows a typical purge connection to the purge tube on the back of<br />

the pump.<br />

To install the nitrogen<br />

(Figure 2D-5)<br />

Attach gas supply by slipping the plastic tube over the<br />

purge connector, see Figure 2D-5.<br />

2D-6