D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong>: Model 1000D<br />

Section 2C 1000D Liquid System Connection & Accessories<br />

CAUTION<br />

The pressure in the wash gland and line should NEVER<br />

exceed the system pressure. A wash pressure greater than the<br />

system pressure may force wash liquid past the pump’s main<br />

seals and contaminate the delivery fluid.<br />

Typically, a small pump delivers the wash fluid to one of the two<br />

1 8" tubes. The second tube would then be used as an outlet to<br />

drain the wash fluid (refer to Figure 2C-3). Select a wash fluid<br />

that will best flush the cylinder of any residue left by the pumped<br />

fluid, yet will not damage the seals. If the wash pump is configured<br />

to recirculate the wash fluid, ensure that you change the<br />

wash fluid at regular intervals.<br />

CAUTION<br />

If the primary pump seal fails, the pressure of the delivery fluid will<br />

be exerted on the secondary (wash gland) seal.<br />

Note<br />

See Page 2A-16 for a list of optional seals.<br />

1/8” Wash Tubes<br />

1/8” Tubing*<br />

Pump*<br />

Outlet Fluid*<br />

Inlet Fluid*<br />

* User-supplied<br />

Figure 2C-3 1000D wash fluid connections<br />

2C-5