D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong>: Model 500D<br />

Section 2B 500D Liquid System Connection & Accessories<br />

Note<br />

The drain tube shown in this photo may appear slightly different<br />

than that on your pump.<br />

Figure 2B-2 Drain tube installation<br />

2B.3 Fluid Connection<br />

Accessories<br />

The optional accessories discussed in this section are used to make<br />

fluid connections from the 500D pump to another apparatus.<br />

When making fluid connections, be sure to use the ferrules provided<br />

in the connectors. Then push the tubing completely into<br />

the connector and finger tighten. Then tighten with a wrench to<br />

clamp the ferrules on the tubing.<br />

2B.3.1 Manual Refill Kit The manual refill valve package is rated to 689.5 bar. It includes<br />

those items necessary to connect to pressurized or non-pressurized<br />

refill sources. Table 2B-2 lists the contents of the<br />

package. Figure 2B-3 illustrates the installation of the package.<br />

To install the refill kit<br />

1. Attach the valve to the pump tower using the valve spacer<br />

and screws provided.<br />

2. Screw the male adapter (P/N 209-0161-01) into the inlet<br />

port of the pump.<br />

3. Connect the pre-bent tubing from the valve to the male<br />

adapter, using the nuts and ferrules supplied.<br />

2B-4