D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong>: Model 500D<br />

Section 2B 500D Liquid System Connection & Accessories<br />

2B.2 Liquid System<br />

Connections<br />

The following section provides general information concerning<br />

the 500D syringe pump, some tips about liquid connections and<br />

information about tubing and tubing cutting procedures. The<br />

accessory package is also discussed.<br />

2B.2.1 Ports There are two ports in the cylinder cap. The 500D ports come<br />

standard with 1 8" pipe thread fittings.<br />

Either port can serve as the inlet or outlet. As shown in Figure<br />

2B-1, you may plug one port and use a single port as both the<br />

inlet and outlet.<br />

DANGER<br />

RISK OF INJURY. THE PRES-<br />

SURE PRODUCED COULD BE<br />

260 BAR. PLEASE UTILIZE<br />

APPROPRIATE TUBING AND<br />

CONNECTIONS NOTED IN THE<br />

MANUAL.<br />

Figure 2B-1 Liquid system plumbing connections for the<br />

500D pump<br />

2B.2.2 500D Installation Tips • Be sure to keep the tubing as straight as possible at the<br />

end, as this will make it easier to install the ferrules.<br />

• Be sure to cut the ends of the tubing squarely.<br />

• Don’t leave burrs on the ends of the tubing.<br />

• When installing ferrules on the tubing, be sure the<br />

tubing extends beyond the ferrule to allow for proper<br />

crimping.<br />

• If the connection is leaking, retighten fittings.<br />

• Push the tubing completely into the port before tightening<br />

the nut.<br />

• When connections are made to the cylinder cap, the<br />

pressure reading may be affected. If the pressure no<br />

longer reads zero, push the pressure zero key on the<br />

controller front panel.<br />

2B.2.3 Tubing Cutting To prevent possible problems, it is important to squarely cut the<br />

tubing. Square ends will be easier to insert through the ferrule<br />

and will lower the dead volume.<br />

2B-2