D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong>: Models 65DM, 100DM, 100DX, 260D<br />

Section 2A 65DM, 100DM, 100DX, and 260D Liquid System Connections & Accessories<br />

ferrule. Connect the other end to the top port on the valve<br />

using the valve fitting.<br />

Note<br />

When nuts are torqued to the cylinder cap ports, the pressure<br />

reading may be affected. If the pressure no longer reads zero,<br />

push pressure zero key on the front panel of the controller to<br />

readjust.<br />

3. Then use the valve fittings to attach the 5.1 cm length of<br />

1 8" OD stainless steel tubing to the port of the 2-way valve.<br />

4. Connect the reducing union, (P/N 209-0162-00) to this 5.1<br />

cm tubing.<br />

5. Connect the 0.9 m length of 1 16" tubing between the reducing<br />

union and your apparatus. Cut to the desired length.<br />

DANGER<br />

RISK OF INJURY. THE PRES-<br />

SURE PRODUCED COULD BE<br />

700 BAR. PLEASE UTILIZE<br />

APPROPRIATE TUBING AND<br />

CONNECTIONS NOTED IN THE<br />

MANUAL.<br />

Figure 2A-6 Outlet valve package connection<br />

2A.3.4 In-Line Filter Package The optional in-line filter package (P/N 68-1247-011) should be<br />

used when it is important to filter flow exiting the pump. This<br />

package (Table 2A-4) contains a 0.5 µm filter and 1 16" tubing.<br />