Section 1 Introduction<br />

9. Press softkey D, PREVIOUS, to return to the main menu.<br />

10. Press softkey D, RETURN, to exit the main menu.<br />

11. Push the blue CONST FLOW key to set the pump mode to<br />

constant flow. CFa will be displayed in the upper left corner<br />

of the screen.<br />

12. Press softkey A, FLOW RATE. The words “ENTER FLOW<br />

RATE” should flash on the display. Use the numeric keys<br />

to enter “1”, “0”, a flow rate of 10 ml/min. Press the ENTER<br />

key to load this setpoint.<br />

Note<br />

If you make an error, press the orange CLEAR ENTRY key to<br />

delete it.<br />

13. Press the blue RUN key. Observe the flow rate displayed<br />

on the first line. After a few moments, the setpoint and<br />

flow rate display should match.<br />

14. Once the setpoint and flow rate match, press the STOP<br />

key.<br />

If more than one pump is connected to the controller, you<br />

will be prompted to press softkey A to stop pump A, B to<br />

stop pump B, etc.<br />

Press softkey A to stop pump A or softkey D to stop all<br />

pumps.<br />

If you encountered any problems during the preliminary<br />

checkout, please contact the Teledyne <strong>Isco</strong> Service<br />

Department. The number is (800) 775-2965 or (402)<br />

464-0231.<br />