D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 1 Introduction<br />

60-1242-306<br />

Figure 1-7 Pump rear panel connectors<br />

Table 1-11 Pump Rear Panel Connectors<br />

Item No. on<br />

Figure 1-7<br />

Connector<br />

Description<br />

1 Pressure outputs Two red binding post/banana jacks providing pressure output voltages.<br />

Pump Left Jack Right Jack<br />

VDC/psi VDC/psi<br />

65D 1.0 V/2000 psi 0.1 V/2000 psi<br />

All others 1.0 V/1000 psi 0.1 V/1000 psi<br />

2 Ground A black binding post/banana jack providing a connection to circuit<br />

common.<br />

3 Pressure transducer The pressure transducer cable must be plugged in for the pump<br />

to operate.<br />

4 Control cable This cable connects the pump to the controller.<br />

5 Mains IEC power connector with EMI filter.<br />

6 Fuses * Limits pump current drawn from main power supply.<br />

Replace with same type: (“T” time delay fuses)<br />

1 - 2.0 Amp for 100/117 volt operation<br />

2 - 1.0 Amp for 234 volt operation<br />

To remove, rotate cap counterclockwise.<br />

7 Chassis ground Ground point for high static installations.<br />

* Fuses not replaceable<br />

by the operator<br />

F101 4.0 Amp “T”<br />

F102, F104 1.5 Amp “T”<br />

F103 0.75 Amp “T”<br />