07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 1 Introduction<br />

POWER REQUIREMENTS a<br />

(Mains voltage line cord is a<br />

“Disconnect Device”)<br />

LINE FREQUENCY<br />

Table 1-5 260D Technical Specifications<br />

100 ± 10 Vac, 1.5 A maximum<br />

117 ± 12 Vac, 1.5 A maximum<br />

234 ± 23 Vac, 0.75 A maximum<br />

50 or 60 Hz<br />

}<br />

Factory Set<br />

LINE VOLTAGE NOISE TOLERANCE<br />

1.7 nominal rms line voltage, 10 µsecond pulses, any phase<br />

angle, random or repetitive<br />

DIMENSIONS PUMP CONTROLLER<br />

Width: 27.18 cm 27.18 cm<br />

Depth: 46.74 cm 30.48 cm<br />

Height: 101.09 cm 13.59 cm<br />

WEIGHT PUMP CONTROLLER<br />

32.8 kg 2.96 kg<br />

FLOW RATE RANGE See Figure 1-2.<br />

FLOW RATE ACCURACY b<br />

FLOW RATE DISPLAY RESOLUTION<br />

ANALOG OUTPUT ACCURACY c<br />

DISPLACEMENT RESOLUTION<br />

REFILL TIME<br />

REFILL OR DEPRESSURIZATION RATE<br />

PRESSURE RANGE<br />

PRESSURE ACCURACY<br />

PRESSURE REPEATABILITY d<br />

ZERO PRESSURE DRIFT<br />

PRESSURE DISPLAY RESOLUTION<br />

± 0.5% (maximum 0.50 µl/min seal leakage)<br />

1.0 µl/min<br />

± 1% of selected range<br />

16.63 nl<br />

2.5 minutes<br />

1.0 µl/min to 107 ml/min at any pressure from 0 to 517.1 bar<br />

0.6895 to 517.1 bar<br />

± 0.5% of full scale at constant temperature<br />

± 0.5% of full scale within 48 hours at constant temperature<br />

± 0.25% of full scale within 48 hours at constant temperature<br />

6.895 kPa<br />

AMBIENT TEMPERATURE RANGE 5 to 40°C<br />

TEMPERATURE DRIFT<br />

HUMIDITY<br />

CYLINDER CAPACITY<br />

DEAD (HEADSPACE) VOLUME e<br />

± 0.15% of full scale/°C<br />

95% maximum<br />

266.05 ml<br />

2.10 ± 0.020 ml<br />

POLLUTION DEGREE 2<br />

INSTALLATION CATEGORY<br />

MAXIMUM ALTITUDE<br />

II<br />

2000 m<br />

a. Underwriters Laboratories (UL) has certified all D <strong>Series</strong> <strong>Pumps</strong> with the exception of the 100 Vac versions.<br />

b. Using water at 137.9 bar and a temperature controlled environment at 30°C.<br />

c. The analog output is an optional accessory.<br />

d. Pressure repeatability specification is based upon re-zeroing pressure transducer every 48 hours. Refer to sub-section<br />

ZERO PRESS in Section 3 of the manual for re-zeroing procedure.<br />

e. Volume in and above the piston seal, head clearance at automatic shutoff, and inlet and outlet ports to the fittings.<br />

1-6

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!